Feature
▪ Automatic zero correction
▪ Peak detection
▪ Data hold
▪ Quick response, and quick return to zero
▪ Power supply by a large-capacity lithium battery, battery level detection, and undervoltage prompt
▪ Two sensors which can switch automatically to detect hydrogen leakage of the ppm level or within the minimum explosion limit
▪ Segment code LCD, and auxiliary alarm threshold LED (10ppm, 100ppm, 1000ppm, and 1%vol)
▪ Audible and visual alarms that have different frequencies in different alarm conditions and can be eliminated manually
▪ Support for user password modifications for sake of user operation protection
Parameter |
Value |
Measured Gas |
Hydrogen (H2) |
Range |
(0~1000) ppm; (0.1~4) %vol |
Maximum Permissible Error |
(0~1000) ppm: ±30 % of the displayed value |
(0.1~4) %vol: ±5 %FS |
|
Response Time |
<2 s |
Resolution |
1 ppm |
Detection Mode |
Pump suction (flow: 200 mL/min) |
Display Mode |
LCD, auxiliary alarm threshold LED |
Battery |
2300 mAh 3.7 V rechargeable lithium battery |
Charger |
Input: (220~240) VAC, 50/60 Hz, 0.3 A |
Output: 5.0 V 0.5 mA |
|
Duration |
6 h |
Operating Temperature |
(-20~+55) ℃ |
Storage Temperature |
(-20~+60) ℃ |
Ambient Humidity |
(20~80) %RH (no condensing) |
Dimensions |
Probing rod length: 320 mm |
169 mm × 60 mm ×28 mm (L×W×H) |
|
Weight |
About 400 g |
Explosion-Proof Sign |
Ex ib ⅡC T3 Gb |